Chloroform Clean Up – Lexie’s QPX RNA from 20110504

After submission of QPX samples to HTGU for Illumina library prep yesterday, I was notified that there was insufficient RNA for the QPX RNA samples. I checked the source RNA on the Roberts Lab NanoDrop1000 and determined that they had high phenol contamination (large peak at 270nm), which results in a large exaggeration in the OD260 absorbance (NanoDrop1000 report[JPEG]; notice terrible OD260/280 ratios; did not save screen shot of absorbance peaks.). As such, the concentrations that Lexie had listed in her notebook for these samples are highly inaccurate and highly inflated. To remove the phenol, I brought all of her QPX RNA samples from 20110504 up to ~200uL with 0.1%DEPC-H2O, added 200uL of chloroform, vortexed for 30s, spun at 12,500g RT for 15mins, and transferred aqueous phase to new tube. Then performed an ethanol precipitation on the aqueous phase. Added 0.1 vols of 3.0M sodium acetate (pH = 5.2), 2.5 vols of 100% EtOH, mixed and incubated at -20C for 1hr. Pelleted RNA by spinning at 16,000g 4C for 15mins.

Results:

As suspected, most of these samples have absolutely no RNA in them. However, the samples that do (the “Control” samples), look great! Pooled 2ug each of the RT Control a & b samples and pooled 2ug each of the 10C Control a & b samples (which are ATCC). Calculations are here. Will take them down to HTGU tomorrow to replace the bad samples that were provided yesterday.

Leave a Reply

Your email address will not be published. Required fields are marked *


e.g. 0000-0002-7299-680X

You may use these HTML tags and attributes: <a href="" title=""> <abbr title=""> <acronym title=""> <b> <blockquote cite=""> <cite> <code> <del datetime=""> <em> <i> <q cite=""> <strike> <strong>