Tag Archives: bioanalyzer

RNA Fragmentation – Herring Liver mRNA for SOLiD Libraries

Samples from 20091203. 0.5uL was removed from each and transferred to separate tubes and diluted to < 5ng/uL for subsequent Bioanalyzer analysis using the Pico chip. Samples were fragmented using RNase III according to the Ambion WTK protocol and then cleaned up/concentrated using the Invitrogen RiboMinus Concentration Module according to the Ambion WTK protocol.

Samples were spec’d prior to running on the Bioanalyzer:

Concentrations/absorbance values are not accurate when using the NanaDrop after using the RiboMinus Concentration module, according to the Ambion WTK protocol. However, yields seem pretty good…

Total, mRNA and fragmented mRNA from each of the four samples was run on the Pico chip with the Eukaryote Total RNA Bioanalyzer protocol.

Results:

The 2L tot (total RNA) and 3L tot (total RNA) samples are clearly very good quality. 2L tot does exhibit some very slight degradation, though. 4L tot (total RNA) and 6L tot (total RNA) show a much greater degree of degradation. All mRNA samples show complete removal of any trace, contaminating rRNA. The fragmented samples (the last four samples on the gel image above) all appear to be perfect. The 4L frag sample simply has less RNA loaded and that is why it is not as dark as the other three fragmented samples. Despite the degradation in the 4L tot and 6L tot samples, the fragmentation profile looks good and we will proceed with making the cDNA libraries for those samples.

Bioanalyzer Submission – Trout RBC, Colleen’s gigas GE sample, Mac’s DH/BB PCR for SOLiD WTK

Samples were delivered for analysis on the DNA 1000 chip.

UPDATE 20091008 They do not have the DNA 1000 kit in stock. Will be using High Sensitivity Kit instead. Will have in stock on Tuesday.

Results:

Bioanalyzer Submission – Rick’s trout RBC samples (various dates)

Submitted Rick’s trout RBC samples to FHRC for bioanalysis using the PicoChip for use with the SOLiD WTK. Submission sheet is here.

Results: Received 20091001.

Lanes 1 & 2 = ribo-depleted AND polyA enriched

Lanes 3 & 4 = ribo-depleted only

Lanes 5 & 6 = total RNA

Lanes 7 & 8 = ribo-depleted only

mRNA – Submission for Agilent Bioanalyzer

Submitted 4uL (108ng/uL) of mRNA from gigas gill (from DATE!; sometime in late August) to the Center for Array Technologies (CART) at UW. Sample was labelled “SR01″.

Results:

Results were received on 20090210. Comment from CAT staff member was that it looked “pretty good.” Here are links to the gel (look for SR01), the electoropherogram of SR01, and the EGRAM of the ladder.