Library Cleanup – LSU C.virginica MBD BS Library

I was contacted by the sequencing facility at the University of Oregon regarding a sample quality issue with our library.  As evidenced by the electropherogram below, there is a great deal of adaptor primer dimer (the peak at 128bp):

 

This is a problem because such a high quantity of adaptor sequence will result in the majority of reads coming off the Illumina being just adaptor sequences.

With the remainder of the library sample prepared earlier, I performed the recommended clean up procedure for removing adaptor sequences in the EpiNext Post-Bisulfite DNA Library Preparation Kit – Illumina (Epigentek).    Briefly:

  • Brought sample volume up to 20uL with NanoPure H2O (added 9.99uL)

  • Added equal volume of MQ Beads

  • Washed beads 3x w/80% EtOH

  • Eluted DNA w/12uL Buffer EB (Qiagen)

After clean up, quantified the sample via fluorescence using the Quant-iT DNA BR Kit (Life Technologies/Invitrogen).  Used 1uL of the sample and the standards.  All standards were run in duplicate and read on a FLx800 plate reader (BioTek).

Results are here: 20150122 – LSU_virginicaMBDlibraryCleanup

Library concentration = 2.46ng/uL

Brought the entire sample up to 20uL with Buffer EB (Qiagen) and a final concentration of 0.1% Tween-20 (required by the sequencing facility).

Sent sample to the University of Oregon to replace our previous submission.